The HI96705 portable photometer is for the measurement of silica. Hanna’s portable photometers feature an advanced optical system; the combination of a special tungsten lamp, a narrow band interference filter, and silicon photodetector ensure accurate photometric readings every time. The Hanna exclusive CAL Check™ feature utilizes ready-made, NIST traceable standards to verify both meter validation and calibration.
Produk asli Kami distributor langsung Hanna Harga murah. Pengiriman ke seluruh Indonesia.
Whatsapp : 0821 3010 0456
Email : [email protected]
CV. Syariftama Multi Global has a wide range of Hanna products – search by type, or model, or filter by brand to find the one you are after. If you still can’t find it contact us..
The HI96705 portable photometer is for the measurement of silica. Hanna’s portable photometers feature an advanced optical system; the combination of a special tungsten lamp, a narrow band interference filter, and silicon photodetector ensure accurate photometric readings every time. The Hanna exclusive CAL Check™ feature utilizes ready-made, NIST traceable standards to verify both meter validation and calibration. The exclusive cuvette locking system ensures that the cuvette is inserted into the measurement cell in the same position every time to maintain a consistent path length.
CAL Check™ – Allows for performance verification and calibration of the meter using a CAL Check NIST traceable secondary standard.
GLP – Records the date of the last user calibration performed.
Built in timer – Allows the use of a timer to ensure the appropriate 3 minutes and 30 seconds reaction time is used for the chemical reaction. At the end of the timer countdown the meter will automatically take the reading. This feature ensures consistency across multiple users.
Indexing indent on meter – The meter has an indent that allows for a lock and key fit with the cuvette cap. This ensures that the cuvette is indexed consistently (same position) in order to maintain the same path length for accurate results.
Cooling Lamp Indicator – It is necessary to maintain a consistent temperature of the optical components in order to maintain a narrow wavelength band of light. This photometer has a cooling lamp indicator that is displayed for a short period of time before each measurement in order to allow the components to cool and obtain the highest accuracy possible.
Error messages – Messages on display alerting to problems including no cap, over range and under range readings, and light source error.
Auto-shut off – Automatic shut off after 10 minutes of non-use when the meter is in measurement mode. Prevents wastage of batteries in the event the meter is accidentally not turned off.
Units of measure – Appropriate unit of measure is displayed along with reading.
Silica is found in all natural waters in the dissolved mineral form. Silica is only slightly soluble in water and can be found as ionic silica, silicates, or colloidal or suspended particles. The solubility of silica is highly dependent on pH, temperature and pressure. Silica’s presence in industrial applications, particularly high pressure turbines, is undesirable because of the scaling caused by the elevated temperature and pressure. Heating systems and reverse osmosis plants also require monitoring of silica to ensure process efficiency. The HI96705 uses an adaptation of the ASTM D859 heteropoly blue method to measure silica concentrations of less than 2.00 mg/L (ppm). When the reagent is added to samples containing silica, the sample will turn a blue hue; the greater the concentration, the deeper the color. The associated color change is then colorimetrically analyzed according to the Beer-Lambert Law. This principle states that light is absorbed by a complementary color, and the emitted radiation is dependent upon concentration. For silica determination, a narrow band interference filter at 610 nm (orange) allows only orange light to be detected by the silicon photodetector and omits all other visible light emitted from the tungsten lamp. As the change in color of the reacted sample increases, absorbance of the specific wavelength of light also increases, while transmittance decreases.
The HI96705-11 CAL Check™standards are used for calibration and performance verification of photometers with the CAL Check function.
SKU | HI96705-web |
---|---|
Product Name | Silica Low Range Portable Photometer – HI96705 |
Quote Required | Yes |
Silica Range | 0.00 to 2.00 mg/L (ppm) |
Silica Resolution | 0.01 mg/L (ppm) |
Silica Accuracy | ±0.03 mg/L ±3% of reading |
Silica Method | adaptation of the ASTM D859 heteropoly blue method |
Photometer/Colorimeter Light Source | tungsten lamp |
Photometer/Colorimeter Light Detector | silicon photocell with narrow band interference filter @ 610 nm |
Battery Type/Life | 9V battery |
Automatic Shut-Off | after ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder |
Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing |
Weight | 360g (12.7oz.) |
Dimensions | 193 x 104 x 69 mm (7.6 x 4.1 x 2.7”) |
Ordering Information | HI96705 is supplied with sample cuvettes (2) with caps, 9V battery, instrument quality certificate and instruction manual. |