Hanna HI-96705 Silica Low Range Portable Photometer

Share on facebook
Share on twitter
Share on linkedin
Hanna HI-96705 Silica Low Range Portable Photometer

Hanna HI-96705 Silica Low Range Portable Photometer

The HI96705 portable photometer is for the measurement of silica. Hanna’s portable photometers feature an advanced optical system; the combination of a special tungsten lamp, a narrow band interference filter, and silicon photodetector ensure accurate photometric readings every time. The Hanna exclusive CAL Check™ feature utilizes ready-made, NIST traceable standards to verify both meter validation and calibration.

 


Produk asli Kami distributor langsung Hanna Harga murah. Pengiriman ke seluruh Indonesia.


Whatsapp :  0821 3010 0456
Email : [email protected]

 

Hanna HI-96705 Silica Low Range Portable Photometer

CV. Syariftama Multi Global has a wide range of Hanna products – search by type, or model, or filter by brand to find the one you are after. If you still can’t find it contact us..

The HI96705 portable photometer is for the measurement of silica. Hanna’s portable photometers feature an advanced optical system; the combination of a special tungsten lamp, a narrow band interference filter, and silicon photodetector ensure accurate photometric readings every time. The Hanna exclusive CAL Check feature utilizes ready-made, NIST traceable standards to verify both meter validation and calibration. The exclusive cuvette locking system ensures that the cuvette is inserted into the measurement cell in the same position every time to maintain a consistent path length.

Features at-a-glance :

CAL Check – Allows for performance verification and calibration of the meter using a CAL Check NIST traceable secondary standard.

GLP – Records the date of the last user calibration performed.

Built in timer – Allows the use of a timer to ensure the appropriate 3 minutes and 30 seconds reaction time is used for the chemical reaction. At the end of the timer countdown the meter will automatically take the reading. This feature ensures consistency across multiple users.

Indexing indent on meter – The meter has an indent that allows for a lock and key fit with the cuvette cap. This ensures that the cuvette is indexed consistently (same position) in order to maintain the same path length for accurate results.

Cooling Lamp Indicator – It is necessary to maintain a consistent temperature of the optical components in order to maintain a narrow wavelength band of light. This photometer has a cooling lamp indicator that is displayed for a short period of time before each measurement in order to allow the components to cool and obtain the highest accuracy possible.

Error messages – Messages on display alerting to problems including no cap, over range and under range readings, and light source error.

Auto-shut off – Automatic shut off after 10 minutes of non-use when the meter is in measurement mode. Prevents wastage of batteries in the event the meter is accidentally not turned off.

Battery status indicator
The instrument has a battery level indicator to display reaming life as follows:
  • 3 lines for 100 % capacity
  • 2 lines for 66 % capacity
  • 1 line for 33 % capacity
  • Battery icon blinks when the capacity is less than 10 %.

Units of measure – Appropriate unit of measure is displayed along with reading.

optical system

Silica is found in all natural waters in the dissolved mineral form. Silica is only slightly soluble in water and can be found as ionic silica, silicates, or colloidal or suspended particles. The solubility of silica is highly dependent on pH, temperature and pressure. Silica’s presence in industrial applications, particularly high pressure turbines, is undesirable because of the scaling caused by the elevated temperature and pressure. Heating systems and reverse osmosis plants also require monitoring of silica to ensure process efficiency.   The HI96705 uses an adaptation of the ASTM D859 heteropoly blue method to measure silica concentrations of less than 2.00 mg/L (ppm). When the reagent is added to samples containing silica, the sample will turn a blue hue; the greater the concentration, the deeper the color. The associated color change is then colorimetrically analyzed according to the Beer-Lambert Law. This principle states that light is absorbed by a complementary color, and the emitted radiation is dependent upon concentration. For silica determination, a narrow band interference filter at 610 nm (orange) allows only orange light to be detected by the silicon photodetector and omits all other visible light emitted from the tungsten lamp. As the change in color of the reacted sample increases, absorbance of the specific wavelength of light also increases, while transmittance decreases.

The HI96705 is available as a kit (HI96705C) which includes

  • Portable Photometer
  • Sample cuvettes (2)
  • Caps
  • CAL Check™ standards with certificate
  • Cuvette cleaning cloth
  • Scissors
  • Rugged carrying case
    • Carrying case has thermoformed insert for meter and accessories

*Reagents ordered separately

cal check standards

CAL Check™ standards with certificate

The HI96705-11 CAL Checkstandards are used for calibration and performance verification of photometers with the CAL Check function.

Supplied with Certificate of Analysis

  • Lot number
  • Expiration date
  • Standard value @ 25 °C
  • Reference meter NIST traceable

Provided storage containers

  • Light tight
  • Protects from accidental breakage

photometer steps

Specifications Hanna HI-96705 Silica Low Range Portable Photometer :

 
SKU HI96705-web
Product Name Silica Low Range Portable Photometer – HI96705
Quote Required Yes
Silica Range 0.00 to 2.00 mg/L (ppm)
Silica Resolution 0.01 mg/L (ppm)
Silica Accuracy ±0.03 mg/L ±3% of reading
Silica Method adaptation of the ASTM D859 heteropoly blue method
Photometer/Colorimeter Light Source tungsten lamp
Photometer/Colorimeter Light Detector silicon photocell with narrow band interference filter @ 610 nm
Battery Type/Life 9V battery
Automatic Shut-Off after ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder
Environment 0 to 50°C (32 to 122°F); RH max 95% non-condensing
Weight 360g (12.7oz.)
Dimensions 193 x 104 x 69 mm (7.6 x 4.1 x 2.7”)
Ordering Information HI96705 is supplied with sample cuvettes (2) with caps, 9V battery, instrument quality certificate and instruction manual.

 

 


Hanna HI-96705 Datasheet