The meter is supplied complete with AMCO-AEPA-1 primary turbidity standards used for calibration and performance verification. The HI-93703-11 meets and exceeds the requirements of the ISO 7027 Method for turbidimetric measurements. Additionally, this meter contains a real-time clock and the capabilities to log measurements and transfer them to a PC using the HI-92000 software.
Produk asli Kami distributor langsung Hanna Harga murah. Pengiriman ke seluruh Indonesia.
Whatsapp : 0821 3010 0456
Email : [email protected]
CV. Syariftama Multi Global has a wide range of Hanna products – search by type, or model, or filter by brand to find the one you are after. If you still can’t find it contact us..
The meter is supplied complete with AMCO-AEPA-1 primary turbidity standards used for calibration and performance verification. The HI-93703-11 meets and exceeds the requirements of the ISO 7027 Method for turbidimetric measurements. Additionally, this meter contains a real-time clock and the capabilities to log measurements and transfer them to a PC using the HI-92000 software.
The HI-93703-11 Portable ISO Compliant Turbidity Meter with Data Logging and PC Connectivity is specially designed for water quality measurements, providing reliable and accurate readings, even within low turbidity ranges. The instrument is based on a state-of-the-art optical system which guarantees accurate results, assures long term stability, and minimises stray light and colour interferences. Periodic calibration with the HI-93703-0, -05, and -10 standards compensates for any variations in intensity of the LED light source. The 25 mm round cuvettes composed of special optical glass guarantee the repeatability of turbidity measurements. HI-93703-11 also includes a real-time clock and log-on-demand capability.
Turbidity is one of the most important parameters used to determine the quality of drinking water. Once considered as a mostly aesthetic characteristic of drinking water, significant evidence exists that controlling turbidity is a competent safeguard against pathogens. In natural water, turbidity measurements are taken to gauge general water quality and its compatibility in applications involving aquatic organisms. The monitoring and treatment or wastewater was once solely based on the control of turbidity. Currently, the measurement of turbidity at the end of the wastewater treatment process is necessary to verify that the values are within regulatory standards.
Turbidity of water is an optical property that causes light to be scattered and absorbed, rather than transmitted. The scattering of light that passes through a liquid is primarily caused by the suspended solids present. The higher the turbidity, the greater the amount of scattered light. Even a very pure liquid will scatter light to a certain degree, as no solution will have zero turbidity.
The ISO standard for turbidity measurement uses infrared wavelength of light that is outside the visible spectrum. The primary benefit of the ISO method over the EPA method is the reduction of colour interference. The EPA method uses a tungsten lamp that produces light that contains all of the visible wavelengths of light that we see as white. A solution that is coloured will be adsorbed by a complementary wavelength of light which will affect the turbidity reading. Since the ISO method is outside the visible wavelength of light the colour in the sample does not interfere with the measurement. The USEPA prefers the tungsten lamp since it provides for better accuracy at low ranges and drinking water should not be coloured. The requirements of an optical turbidity meter for the measurement of diffuse radiation used in low range measurements (i.e. drinking water) expressed as formazin nephelometric units (FNU) according to ISO 7027:1999 are:
The light beam that passes through the sample is scattered in all directions. The intensity and pattern of the scattered light is affected by many variables, such as wavelength of the incident light, particle size and shape, refractive index, and colour. The optical system of the HI-93703-11 includes a LED, and a scattered light detector (90°).
The lower detection limit of a turbidimeter is determined by stray light that is detected by the sensors but not caused by light scattering from suspended particles. The optical system of the HI-93703-11 is designed to have very low stray light, providing highly accurate results for low turbidity samples.
The AMCO AEPA-1 standards for HI-93703 ensure that measurements are traceable to a primary reference materials. These standards are used for calibration and performance verification of the turbidity meter.
SKU | HI93703-11 |
---|---|
Product Name | Portable ISO Turbidity Meter with RS232 – HI93703-11 |
Quote Required | Yes |
Turbidity Range | 0.00 to 50.00 FTU |
Turbidity Resolution | 0.01 (0.00 to 50.00 FTU); 1 (50 to 1000 FTU) |
Turbidity Accuracy | ±0.5 FTU or ±5% of reading (whichever is greater) |
Turbidity Calibration | three points (0 FTU, 10 FTU and 500 FTU) |
Turbidity Light Source | silicon photocell |
Turbidity Light Detector | infrared LED / life of instrument |
Logging Type | log on demand |
Logging Memory | 199 measurements |
Connectivity | RS232 using HI92000 software |
Automatic Shut-Off | auto-off after 5 minutes of non-use |
Battery Type/Life | 1.5V AA (4) /approximately 60 hours of continuous use or 900 measurements |
Environment | 0 to 50°C (32 to 122°F); RH max 95% (non condensing) |
Dimensions | 220 x 82 x 66 mm (8.7 x 3.2 x 2.6’’) |
Weight | 510 g (1.1 lb.) |
Ordering Information | HI93703-11 is supplied complete with a glass cuvette, batteries, and instructions. |